MPV is possible with a HMV with existing MPV option and/or most traditional volume-cycled
pressure/flow-triggered home ventilators. The set up configuration will differ depending
on the type of ventilator available.
Bilevel pressure devices and home ventilators in pressure modes are NOT appropriate
for MPV. The inspiratory positive airway pressure (IPAP or pressure support) setting,
limits LVR and cough efficacy; the expiratory positive airway pressure (EPAP or
PEEP) generates a constant flow-by which cannot be minimized to 0 cmH2O and may
be a nuisance to user and; in some instance, the device low EPAP alarm cannot be
disengaged, enabling a nuisance alarm during MPV.
Clinical Considerations
- A learning period may be necessary for individuals who have been ventilated with
an artificial airway for an extended period or who have not used LVR with a modified
resuscitation bag (e.g., breath-stacking with LVR bag).
- The individual needs to generate very little effort via the mouthpiece to initiate
a ventilator delivered breath.
- With lips placed firmly around the mouthpiece, the individual uses their cheek muscles
to draw or “sip” through the mouthpiece, triggering a ventilator-assisted breath
that provides as much air volume as needed.
- The individual must be able to close the soft palate, seal the nasopharynx, and
open the glottis and vocal cords to allow the ventilator delivered breath to reach
the lungs.
- Instruct the individual to accept as much air as possible in to the lungs up to
the preset pressure limit. Average high pressure limit ranges from 45 to 70 cmH2O
to allow for MIC.
- The individual should feel a satisfying breath volume with every inspiratory effort.
The inspiratory volume, flow, time, and rise settings should be adjusted to promote
individual comfort.
- Exhalation should occur naturally through the nose or open mouth.
- A filter proximal to the mouthpiece will soften the ventilator circuit constant
flow-by and may improve individual comfort.
- A one-way valve proximal to the mouthpiece may be necessary if the individual is
unable to exhale through the nose or at the side of the mouth. The one-way valve
will alleviate ventilator alarms caused by back-flow (individual exhalation) to
the circuit.
- Once the individual is familiar with taking ventilator supported breaths via the
mouthpiece, they can start to learn LVR by breath-stacking. This requires taking
multiple MPV breaths without exhaling until the lungs are full. LVR with MPV should
be followed by gentle exhalation, a breath-hold up to 5 seconds or individual initiated
cough effort.
- LVR with MPV should not induce dizziness or chest discomfort.
- Individual and clinician communication is essential when determining alarm settings
to ensure safe, comfortable and effective ventilation.
LVR with MPV is recommended as often as required; with a minimum of 2-3 times per
day with 3-5 lung stretches per session. If combined with manually assisted cough
(MAC) it is best done before meals and at bedtime to minimize risk of refluxed gastric
content. MPV is possible with most home mechanical ventilators (HMV). In the following
section we describe the how to set up MPV with the following devices and related
accessories:
A) HMV with existing MPV option e.g., Trilogy 200; and
B) Traditional volume-cycled pressure/flow-triggered HMV with, and without, a proprietary
circuit.
A) MPV with the Trilogy 200 HMV
Parameter settings recommendations for MPV with a volume-cycled pressure/flow-triggered
HMV are:
Parameter Settings
|
Recommendations
|
Circuit Type
|
Passive
|
Mode
|
Assist / Control
|
MPV
|
ON
|
Tidal Volume (VT)
|
larger than spontaneous VT enabling LVR to MIC within 2 to 3 stacked breaths
|
Breath Rate
|
0 if the individual has sufficient ventilator-free breathing time (VFBT) or up
to 12 if ventilator dependent
|
Inspiratory Time
|
1.2 to 1.5 seconds; adjust for comfort and desired peak inspiratory flow (PIF);
PIF will be dependent on VT setting
|
Flow Pattern
|
Ramp or square as per individual comfort
|
PEEP
|
0 cmH2O
|
Low Inspiratory Pressure
|
1 to 2 cmH2O
|
High Inspiratory Pressure
|
Up to 70 cmH2O to allow for LVR to MIC
|
Apnea or Circuit Disconnect
|
MUST be enabled if the individual has limited VFBT or if close monitoring is required
|
Other
|
Refer to the clinical manual and software version for further parameters applicable
to MPV
|
B) MPV with a traditional volume-cycled pressure/flow-triggered HMV
Minimal requirements for MPV with volume-cycled pressure/flow-triggered HMV include:
- Assist / Control mode;
- A fixedinspiratory time setting that is notdependent on flow and I:E ratio settings;
- Ability to set positive end expiratory pressure (PEEP) low limit capability at 0
cmH2O;
- Apnea and low breath rate alarms that can be turned OFF; and
- Single-limb open-circuit with flow limiting mouthpiece assembly to allow the circuit
to remain open to air without triggering the low pressure or apnea alarms.
Parameter settings recommendations for MPV with a volume-cycled pressure/flow-triggered
HMV are:
Parameter Settings
|
Recommendations
|
Mode
|
Assist Control; pressure modes are NOT appropriate for MPV
|
Tidal Volume (VT)
|
SetVT larger than spontaneous VT enabling LVR to MIC
within 2 to 3 stacked breaths
|
Breath Rate
|
0 if the individual has sufficient VFBT or up to 12 if ventilator dependent
|
Inspiratory Time
|
1.2 to 1.5 seconds; adjust for comfort and desired PIF; PIF will be dependent
on VT setting
|
Rise or Flow
|
Adjust as per individual comfort
|
Sensitivity
|
Adjust to enable minimal individual effort yet preventing auto-cycling
|
PEEP
|
0 cmH2O
|
Low Inspiratory Pressure
|
1 to 2 cmH2O; minimal setting to create sufficient back-pressure against
the flow limiting mouthpiece or flow-restrictor
|
High Inspiratory Pressure
|
Up to 70 cmH2O to allow for LVR to MIC
|
Apnea or Circuit Disconnect
|
MUST be enabled if the individual has limited VFBT or if close monitoring is required;
Set the low breath rate alarm one breath above the ventilator set breath rate
|
Other
|
Refer to the clinical manual and software version for further parameters applicable
to MPV
|
MPV accessories applicable to HMVs are described in the table hereunder followed
by a brief item description.
MPV Accessories
|
Trilogy 200 HMV
|
Volume-cycle pressure/flow-triggered HMV with a circuit requirement that is:
|
non-proprietary
|
proprietary
|
Flexible tapered single-limb circuit
|
√
|
√
Use the flexible tapered single-limb circuit in place of the usual HMV non-proprietary
circuit; the exhalation valve is not required.
|
√
Add the flexible tapered single-limb circuit beyond the proprietary circuit
Y connector or exhalation valve.
|
Loc-Line modular circuit-support arm with clamp
|
√
|
√
|
√
|
Pressure port connector for pressure line
|
Not required
|
√
Place the circuit pressure port connector proximal to the mouthpiece for inspiratory
trigger sensitivity.
|
√
It is best to use the existing pressure port within the proprietary circuit assembly
|
Mouthpiece assembly
|
√
|
√
|
√
|
Flow-restrictor
|
Not required
|
Not required
|
Not required
|
Bacteria filter
|
√
|
√
|
√
|
Accessories Description
- Flexible tapered single-limb circuit; 3-6 feet in length; 22 mm
I.D. at the ventilator end and 15 mm I.D. at the user end; this type of circuit
performs well when using the Loc-Line modular circuit-support arm which stabilizes
circuit placement and facilitates ease of access to the mouthpiece.
- Loc-Line modular circuit-support arm with clamp; to assemble the
circuit and support arm, remove the clamp, roll the Loc-Line arm on a flat surface
until it is straight, then push the 15 mm I.D. circuit end through the clamp end;
keep pushing the circuit until it reaches half an inch beyond the other end of the
Loc-Line. The Loc-Line circuit-support arm can be secured to a wheelchair cane,
or table top using the clamp.
- Mouthpiece assembly appropriate for the individual’s needs e.g.,
adult or pediatric mouthpiece or tapered nipple connector with Tygon® tubing, with
or without an elbow connector.
- Flow-restrictor may be necessary when using the adult mouthpiece
as it will allow the ventilator circuit to remain open to air without activating
the low pressure or apnea alarm.
- Bacteria filter proximal to ventilator or proximal to mouthpiece.